Specification: Part Number 1071083 Supply voltage: 12 V DC ... 30 V DC Resolution: < 2 mm Lower signal level: 3.8 mA ... 4 mA Output current: < 100 mA Probe type: Mono rod probe Length coaxial cable: 2 m, PVC Output load: 0 V ... 10 V > 750 Ohm at Uv >= 14 V, 4 mA ... 20 mA < 350 Ohm at Uv > 12 V, 4 mA ... 20 mA < 500 Ohm at Uv > 15 V Inductive load: < 1 H CULus certificate: + Accuracy of sensor element: ± 5 mm Wetted parts: 1.4404, PTFE Signal voltage HIGH: Vs - 2 V Signal voltage LOW: ≤ 2 V Probe length: 2,000 mm Inactive area at process connector: 25 mm Output signal: 1 PNP transistor output (Q1) and 1 PNP / NPN transistor output (Q2) switchable, 4 mA ... 20 mA, 0 V ... 10 V automatic switching depending on the load. MTTF: 194.3 a (EN ISO 13849-1) Capacitive load: 100 nF Hysteresis: Min. 2 mm, free adjustable Ambient storage temperature: -40 °C ... +80 °C Initialization time: ≤ 5 s Upper signal level: 20 mA ... 20.5 mA Connection type: Round connector M12 x 1, 5-pin Response time: < 400 ms Maximum level change: 500 mm/s Dielectricity constant: ≥ 1.8 with coaxial tube, ≥ 5 for mono rod probe / rope probe Power consumption: ≤ 100 mA at 24 V DC without output load Measurement: Switch, Continuous Housing material: Plastic PBT Protection class: III Repeatability: ≤ 2 mm Process connection: G ¾ A Ambient temperature coaxial cable: -20 °C ... +60 °C Max.probe load: 6 Nm Inactive area at probe end: 10 mm Temperature drift: < 0.1 mm/K Process temperature: -20 °C ... +100 °C Medium: Fluids Ambient temperature operation: -20 °C ... +60 °C RoHS certificate: + IO-Link: + EMC: EN 61326-1:2006, 2004/108/EG GOST approval: + Design: Remote amplifier, cable length 2 m Conductivity: No limitation Enclosure rating: IP 67: EN 60529 Process pressure: -1 bar ... 10 bar